Articles
Table 1 Linear regression coefficients of T2 versus temperature (ms/°C) for the different pulse sequence parameters
From:
T2-based temperature monitoring in bone marrow for MR-guided focused ultrasound
Heating
Cooling
Water Suppr. TE = 28/147
7.05
5.39
Water Suppr. TE = 35/182
7.00
5.39
No Water Suppr. TE = 35/182
5.48
4.47