Table 5 Experimental conditions related to the influence of neighboring pulses

From: Simulation of high-intensity focused ultrasound lesions in presence of boiling

  Median cooling Spacing along Spacing along Overlap
  time (s) x (mm) y (mm) (%)
4-s pulses 22 1.8 1.6 45
8-s pulses 39 4 3.5 19
12-s pulses 48 5 4.3 6.7
  1. The overlap is calculated based on elliptic representations of the lesions in the xy plane. It is calculated as the ratio between the area which is covered by at least two individual lesions and the area covered by the global lesion